IMPLEN

89-6646-26 NanoPhotometer NP80 Microspectrophotometer (with LCD touch screen) NP80-TOUCH

기능

  • Overview: [Sample Compression Technology] [True Path Technology] enables extremely simple and accurate measurements without the need for dilution.
  • [Sample Compression Technology]: The unique technology of sandwiching the sample between the optical path window and mirror enables reliable measurement without depending on the surface tension of the sample. High concentration and high reproducibility measurement can be performed with a trace sample of at least 0.3uL. It also supports the measurement of protein solutions containing surfactants and high concentration buffers. The evaporation prevention structure also supports the measurement of volatile samples.
  • True Path Technology: Automatically selects the optimum optical path length based on the absorbance of the sample. Two high-precision optical paths (Dilution rate: 1/15, 1/140) cover a wide range of dynamic ranges.
  • High-end model equipped with all micromeasurement, general-purpose cuvette and vortex.
  • It has an LCD touch screen and can be used standalone.

사양

  • trace/cuvette measurement model
  • < General Specifications >
  • External Dimensions: 200 (W) x 290 (D) x 200 (H) mm (Including Touch Screen)
  • Weight: 3.8~5.2 kg (depending on configuration)
  • Power: 90~250 V, 50/60Hz, 60 W
  • Display: 1024 x 600 pixels
  • Certification: CE, IEC 61010-1:2012, EN 61326-1:2013
  • In/Out Ports: USB A x 2, USB B, HDMI, Ethernet, Wireless LAN
  • Other data input: USB mouse, USB keyboard
  • security: Kensington slots
  • optical system specification
  • Wavelength Scan Range: 200~900 nm
  • Measurement time: 3.5~6.0 seconds
  • Wavelength reproducibility: ±0.2 nm
  • Wavelength accuracy: ±0.75 nm
  • Bandwidth: 1.8 nm or more
  • Stray light: <0.5% 240 nm when using Nal, <1% 280 nm when using acetone
  • Measurement repeatability: <0.002 A (optical path length 0.67 mm) when using 280 nm
  • Accuracy of measurement: <1.75% 0.7 A (optical path length 0.67 mm) when using 280 nm
  • 0 stability: +/- 0.003 A/hour (20 minutes after warm-up when using 280 nm)
  • Noise: 0.002 A rms(0 A when using 280 nm), 0.002 A (pk-pk) (0 A when using 280 nm) 
  • Optics: 1 x 3648 CCD Array
  • Light source: xenon flash lamp
  • Light source life: 10^9 flashes or more than 10 years
  • < During micromeasurement >
  • Detection range/dsDNA: 1 ng/μL to 16,500 ng/μL
  • Detection range/BSA: 0.03 mg/mL to 478 mg/mL
  • Minimum sample volume: 0.3 μL
  • Measuring range: 0.02~330 A
  • Optical path length: 0.67, 0.07 mm
  • Dilution rate: 1/15, 1/140
  • Vortex mixer: 2800rpm
  • < During cuvette measurement >
  • Detection range/dsDNA: 0.1 ng/μL to 130 ng/μL
  • Detection range/BSA: 0.003 mg/mL to 3.7 mg/mL
  • Measuring range: 0~2.6 A
  • Optical path center height (Z): 8.5 mm
  • Cell Type: Outer Dimensions 12.5 x 12.5 mm
  • Temperature control: 37°C±0.5°C
  •  
주문 번호 89-6646-26
모델 번호 NP80-TOUCH
표준 가격 JPY: 2,180,000 USD: 13,563.96
Excange rate 1USD= 160.72JPY
Valid price in Japan
수량 1unit
일본의 주식
공급자 재고

다양한 상품군(多樣한) (다른 사이즈,다른 사양,옵션등)

제품 이미지 주문 번호 이름 모델 번호 수량 표준 가격 일본에서의 유효 가격 재고
[공급자 재고]
89-6646-26 NanoPhotometer NP80 Microspectrophotometer (with LCD touch screen) NP80-TOUCH NP80-TOUCH 1unit JPY: 2,180,000 USD: 13,563.96

89-6646-27 NanoPhotometer N60 Microspectrophotometer (with LCD touch screen) N60-TOUCH N60-TOUCH 1unit JPY: 1,850,000 USD: 11,510.70

89-6646-28 NanoPhotometer N50 Microspectrophotometer (with LCD touch screen) N50-TOUCH N50-TOUCH 1unit JPY: 1,700,000 USD: 10,577.40

89-6646-29 NanoPhotometer C40 spectrophotometer with LCD touch screen C40-TOUCH C40-TOUCH 1unit JPY: 1,100,000 USD: 6,844.20

제품 카탈로그

카탈로그 이름 페이지
AS ONE Catalog 2025 [Instruments for Laboratory] 1324