63-1392-97 Wide range gauge CC-10-ICF70
Fitur
- Covers the area of atmospheric pressure sensor, rough gauge and high vacuum gauge. This unit can measure from atmosphere to ultra-high vacuum without switching.
- It is a transducer type with integrated sensor and circuit, and the pressure display is attached to the body, so measurement can be performed only with the body without external controller.
- No operation required: Simply connect 24 VDC and the pressure is displayed digitally. Automatic measurement from atmospheric pressure to ultra-high vacuum without any operation.
- Equipped with analog output, RS485 interface and 3 set points as standard.
- Tough and safe: Cold cathode gauge is stronger than B-A gauge for high vacuum measurement. Since there is no high-temperature filament, there is no risk of measurement stoppage due to sudden wire breakage.
- Superior accuracy and stability: Crystal gauges are used to measure atmospheric to roughing pressures, which are much more stable than Pirani gauges.
- Easy maintenance: Gauge cleaning, replacement, and other maintenance can be easily performed by the customer.
- Accessories, expandability: AC adapter, dedicated data logger software, external display are available as accessories.
Spec
- Mounting Flange: ICF70
- Pressure measurement display: 1.0 x 10-7 Pa to atmospheric pressure
- Measurement Method: Crystal Vacuum Sensor (Atmospheric Pressure and Roughing Pressure) + Double Inverted Magnetron Cold Cathode Gauge (High Vacuum Side)
- Pressure display: 7-segment LED digital display is standard on the main unit.
- Pressure Signal External Output: Analog Output 0-10 V (Log or Combined Linear Output Selectable)
- Set Points: 3 (with Hysteresis) Standard Equipment 50 VDC, 0.12 A Photomoss Relay 1 point is independent, 2 points are common
- Relay I/O: input for crystal 0 adjustment, atmospheric pressure adjustment, and CCG switching prohibition
- Digital interface: RS485 standard
- Input voltage: 22 DC -26 V
- Power consumption: Max 8W
- Operating Environment: -40°C~+55°C (Storage)
- Ambient temperature: max50°C (operating)
- Baking temperature: 150°C when the control circuit is removed and the sensor is used alone.
- Connection connector: Dsub15 pin (female connector attached)
- Weight: 1060 g
- ICF70 (Rotation) with Flange
- D Sub-connector (female) attached
Ukuran paket:230×105×100 mm 1.19 kg [Tentang Ukuran Paket]
| Perintah No. | 63-1392-97 | |
|---|---|---|
| Model No. | CC-10-ICF70 | |
| Harga standar |
JPY: 303,000
USD: 1,900.64
Excange rate 1USD= 159.42JPY
Valid price in Japan |
|
| Kuantitas | 1piece | |
| Saham di Jepang |
|
|
| Saham Pemasok |
|
|
Variasi produk (ukuran berbeda, spesifikasi berbeda, produk opsional, dll)
| Gambar produk | Perintah No. | Nama | Model No. | Kuantitas | Harga standar | Harga valid di Jepang | Stok [Saham Pemasok] |
|
|---|---|---|---|---|---|---|---|---|
![]() |
63-1392-96 | Wide range gauge CC-10-NW25 | CC-10-NW25 | 1piece | JPY: 288,000 | USD: 1,806.55 |
|
|
![]() |
63-1392-97 | Wide range gauge CC-10-ICF70 | CC-10-ICF70 | 1piece | JPY: 303,000 | USD: 1,900.64 |
|