65-0570-71 Uniace Bath with Low Temperature and Constant Temperature Water Tank 9.3/11 L/min NCB-1220B
Features
- Wide range of temperature, can be controlled and recorded by a personal computer
- Small tabletop type, excellent cost performance
- Jacket column, constant temperature of electrophoresis tank, useful for reaction of micro tube in built-in water tank.
- The temperature range is -20~95°C. It can be used for a variety of applications, including chemical reactions, reagent adjustment, analytical equipment, viscometers, and other external circulation.
- The NCB-1220B type has a narrow pipe diameter and enables stable flow circulation even in the presence of high pressure loss channels, heat exchangers and head differences. It can also circulate a -20°C nib line ® Z-1 model (70wt%).
- It is small and does not take up much space. It can be easily installed on and under the laboratory table.
- When communicating with a personal computer, you can perform remote operations such as constant operation and step operation, and control and record temperature trend data via dedicated software. (A dedicated communication cable is required.)
- The optional external temperature sensor can be used to control the temperature corresponding to the external device and the sample temperature.
- NCB-1220B is a RoHS Directive compliant product. Low GWP product filled with refrigerant (CFC gas) R452A.
Spec
- Performance:
- Temperature adjustment range: - 20 ~ 95 degrees C
- Temperature control accuracy: below +/- 0.1 degrees C
- Cooling capacity (at 10 degrees Celsius): 160 W (137 kcal/h)
- External Circulation Capacity (50/60 Hz) Maximum Flow: 9.3/11 L/min, Maximum Pump Head: 4.3/6.3 m
- Features:
- Temperature control: P.I.D control, contactless zero-cross output
- Temperature setting/display: Sheet key input/digital display, minimum digit 0.1 degrees C
- Safety functions: Upper and lower temperature anomaly detection, water level drop, overshoot, control anomaly, sensor anomaly, high pressure switch, refrigerator overload relay, variable independent overrise detector, leakage and overcurrent breaker
- Timer program function: Auto start, Auto stop step mode
- Included features: Auto-tuning, temperature display correction, power outage return settings
- External input/output: PC communication (RS -232 C synchronous serial communication), external temperature sensor terminal, recorder output terminal, alarm output terminal (Voltage Free Contact (a Contact) Output AC 250 V/Max. 3 A)
- Configuration:
- Heater: 500 W (SUS 316 L)
- Refrigerator/Refrigerant: Air-cooled output: 300 W/R 452 A
- Tank Stirring: Jet Stirring
- Cooling Coil: SUS 316 L
- Tank Lid: SUS 304
- Temperature sensor: Platinum temperature resistor Pt 100 ohm
- Standard:
- Outboard Circulation Nozzle: Both Outlet and Return Outlets 11 mm Outside Diameter Hose Outlet
- Recorder output range: 1 ~ 4 VDC (- 40 ~ 110 deg. C)
- Water tank material: SUS 304
- Tank Inside Dimensions (mm), Capacity: 140 W x 260 D x 150 H, Approx. 4 L
- Effective dimensions in tank (mm): 110 W x 80 D x 130 H
- Operating ambient temperature range: 5 ~ 35 degrees C
- External dimensions (mm): 210 (225) W x 433 (530) D x 600 H
- Mass: About 34 kg
- Power input/voltage: 12 A 1.2 kVA/100 VAC 50/60 Hz
- *Performance is measured at room temperature of 20 degrees Celsius, rated supply voltage, 50 Hz and no load.
- *Use a low-temperature heating medium when the liquid temperature is set below 10 degrees Celsius.
- *It can not be used in conjunction with a load (recirculating heat source) at liquid temperatures above 40 degrees Celsius.
- *Depending on the conditions, the refrigerator may be turned ON/OFF around the set temperature of 40 ~ 65 degrees Celsius, which may reduce the temperature adjustment accuracy. (plus or minus 1.5 degrees Celsius)
- *The temperature adjustment range is - 10 ~ 95 degrees Celsius when the operating ambient temperature is 30 degrees Celsius or more.
- *Dimensions in parentheses include projections.
- Code: 279530
Package size:570×330×690 mm 35 kg [About Package size]
| Order No. | 65-0570-71 | |
|---|---|---|
| Model No. | NCB-1220B | |
| JAN Code | 4573528557142 | |
| Standard price |
JPY: 328,000
USD: 2,056.04
Excange rate 1USD= 159.53JPY
Valid price in Japan |
|
| Quantity | 1unit | |
| Stock in Japan |
|
|
| Supplier Stock |
|
|
Product Variations (Different Sizes, Specifications, Optional Products, etc.)
| Product image | Order No. | Name | Model No. | Quantity | Standard price | Valid price in Japan | Stock [Supplier Stock] |
|
|---|---|---|---|---|---|---|---|---|
![]() |
65-0570-70 | Uniace Bath with Low Temperature and Constant Temperature Water Tank 4.5/5.0 L/min NCB-1220A | NCB-1220A |
|
1unit | JPY: 275,000 | USD: 1,723.81 |
|
![]() |
65-0570-71 | Uniace Bath with Low Temperature and Constant Temperature Water Tank 9.3/11 L/min NCB-1220B | NCB-1220B | 1unit | JPY: 328,000 | USD: 2,056.04 |
|