63-1396-47 Cooling Water Circulation System Coolace CA-1116A
Features
- Standard chillers boasting substantial cooling capacity, with a wide temperature setting range of -20~30 degrees Celsius, for a wide range of applications. The cooling capacity is 1200 W (approximately 1030kcal/h) at a liquid temperature of 10 degrees Celsius (room temperature of 20 degrees Celsius).
- This product complies with the PoHS Directive.
Spec
- Cooling/Circulation System: Air Cooled/Sealed System Directional Circulation
- Temperature setting range: - 20 ~ 30 C (without heater)
- Temperature control accuracy: +/- 2 degrees C
- Cooling capacity (at liquid temperature): 10 degrees C/1200 W (1030 kcal/h), 0 degrees C/950 W (820 kcal/h), -10 degrees C/650 W (560 kcal/h)
- External circulation capacity (50/60 Hz): maximum flow rate 12/14 L/min, maximum head 9.5/13 m (discharge pressure: 0.09/0.13 MPa)
- Temperature control: Refrigerator ON-OFF control
- Temperature setting/display: Sheet key input/digital display, minimum digit 1 degree C
- Safety functions: Leakage/overcurrent breaker, refrigerator high-pressure switch, overload relay holding circuit, refrigerator protection timer, fuse for service outlet, temperature controller self-diagnostic function, circulating pump thermal protector
- Accessory function: With flow stop valve
- External input/output: Communication terminal for vacuum controller (NVC -3000 type), communication terminal for evaporator (N -1210 B type)
- Refrigerator/Refrigerant: Air-cooled output: 650 W/R 407 C
- Cooling Coil: SUS 316 L
- External circulation nozzle: One-touch connector inner diameter 10 mm, one-touch hose nozzle outer diameter 10 mm
- Water Tank Materials, Dimensions, Capacity: SUS 304, Inner Diameter 280 x 270 mm, 16.5 L (Liquid Amount 14 L)
- Service Outlet: 100 VAC for Aspirator, Max. 4 A1 Unit
- Operating ambient temperature range: 5 ~ 35 degrees C
- External dimensions: 340 (354) x 384 x 851 (940)/* Dimensions in () include projections.
- Weight: 41 kg
- Power input/voltage: 11 A/1.1 kVA/100 VAC 50/60 Hz
- *Performance is measured at room temperature of 20 degrees Celsius, rated supply voltage and no load.
- *Cooling capacity is +/- 10% of display capacity.
- *The external circulation capacity is +/- 10% of the display capacity.
- *At room temperature of 35 degrees Celsius, the temperature to reach cooling is -10 degrees Celsius.
- *Pure water can not be circulated.
- Code: 268440
*This product might require extra fees.
Package size:490×450×995 mm 44 kg [About Package size]
| Order No. | 63-1396-47 | |
|---|---|---|
| Model No. | CA-1116A | |
| JAN Code | 4573528553649 | |
| Standard price |
JPY: 268,000
USD: 1,679.94
Excange rate 1USD= 159.53JPY
Valid price in Japan |
|
| Quantity | 1unit | |
| Stock in Japan |
|
|
| Supplier Stock |
|
|
Product Variations (Different Sizes, Specifications, Optional Products, etc.)
| Product image | Order No. | Name | Model No. | Quantity | Standard price | Valid price in Japan | Stock [Supplier Stock] |
|
|---|---|---|---|---|---|---|---|---|
![]() |
63-1396-47 | Cooling Water Circulation System Coolace CA-1116A | CA-1116A |
|
1unit | JPY: 268,000 | USD: 1,679.94 |
|
![]() |
63-1396-48 | Cooling Water Circulation System Coolace CA-2610 | CA-2610 |
|
1unit | JPY: 640,000 | USD: 4,011.79 |
|
Product Catalog
| Catalog Name | Page |
|---|---|
| AS ONE Catalog 2025 [Instruments for Laboratory] | 94 |
| AS ONE Catalog 2023 [Instruments for Laboratory] | 105 |