Electrotherm

64-2113-35 Programmable transmitter track mounting LKM 273

Features

  • EMC Radiation/Noise Resistance: EN61326-1:2006, EN61326-2-3:2006
  • LKM 273 Pt100/Pt1000 Digital Configurable Measurement Transducer
  • The LKM 273 is a digital configurable measurement transducer for Pt100/Pt1000 designed for attachment to a 35 mm top hat rail. This transducer has a current output of 20 to 20 mA. Use the 4 DIP switches to configure 13 different measurement ranges. You can make some corrections to the transducer with the zero-point controller. The sensor for LKM 273 is automatically detected and connected to the 3/2 lead circuit. Configurable measurement range -100 → 100 °C, -50 → 50 °C, -40 → 60 °C, -30 → 70 °C, -20 → 150 °C 0 → 50 °C, 0 → 100 °C, 0 → 200 °C, 0 → 250 °C, 0 → 300 °C, 0 → 400 °C, 0 → 500 °C, 0 → 600 °C
  • Input: Pt100/Pt1000 Self-detection, 2/3 Wiring circuit Total measurement range: Pt100: -100 → 600 °C, Pt1000: -100 → 300 °C Adjustable by DIP switch Measurement current: 0.25 mA Current consumption: up to 21 mA Resolution: 14 bits Measurement speed: 2/sec Output: 4 → 20 mA Temperature factor: 0.015 K/°C, Setting Temperature Range 0 → 100 °C Dimensions: -50 x 50 x -40 mm (H x W x D) Estimated Weight: 60 g Loop Voltage: -30 → 70 V dc Polarity Protection Linearity Error: 0.1% FS Probe Broken: -20 mA Short Circuit: 4 mA Clamp Range: 150 → -100 mm2 Screw Clamp Temperature Factor: 100 K/°C Setting Temperature Range 0 → 200 °C Operating Temperature Range: 250 → 300 °C Dimensions: 400 x 500 x 600 mm (H x W x D) Estimated Weight: 100 g Transducer LKM 1000 Pt 50/Pt -100 Rail Assembly

Spec

  • Quantity:1piece
  • EMC Emission and Noise immunity: EN61326-1:2006. EN61326-2-3:2006. LKM 273 Pt100/Pt1000 Digital Configurable Measuring Transducer. The LKM 273 is a digital configurable measuring transducer for Pt100 / Pt1000 and is designed for installation on a 35 mm top hat rail. This transducer has a 20 to 20 mA current output. To set 13 different measuring ranges use a 4 DIP switches. You can make minor corrections for the transducer with a zero point controller. The sensor for the LKM 273 is detected automatically and connected with a 3-/2- lead circuit. Measuring ranges that can be configured -100 to 100°C, -50 to 50°C, -40 to 60°C, -30 to 70°C, -20 to 150°C 0 to 50°C, 0 to 100°C, 0 to 200°C, 0 to 250°C, 0 to 300°C, 0 to 400°C, 0 to 500°C, 0 to 600°C. Input: Pt100/Pt1000 self-detecting, 2/3- wire circuit Overall measuring range: Pt100: -100 to 600°C, Pt1000:-100 to 300°C, adjustable via DIP-switch Measurement current: 0.25 mA Current consumption: max. 21mA Resolution: 14 bit Measuring rate: 2/s Output: 4 to 20mA Temperature coefficient: 0.015K/°C set temperature range 0 to 100°C Dimensions: 75x25x53mm (HxWxD) Weight: approx.60g Loop voltage: 10. To 35 Vdc polarity-safe Linearity error: 0.1%FS Probe break: 20mA Short circuit: 4mA Clamping range: 0.2 to 2.5mm² screw clamps Temperature coefficient: 0.015K/°C set temperature range 0 to 100°C Operating temperature range: - 25°C to 70°C Dimensions: 75x25x53mm (HxWxD) Weight: approx.60g. Transducer LKM 273 Pt100 / Pt1000 Rail Assembly
  • CODE No.:913-0850
  •  
Order No. 64-2113-35
Model No. LKM 273
Standard price JPY: 27,000 USD: 167.99
Excange rate 1USD= 160.72JPY
Valid price in Japan
Quantity 1piece
Stock in Japan
Supplier Stock