Cytiva (Whatman)

62-6178-68 Poly Cap TC36(With Filling Bell, Pore Diameter 0.2/0.2 μm Sterilized) 6715-3602

Features

  • The filter with 2 layers of PES prevents clogging.
  • Low protein adsorption is recommended for filtering aqueous solutions.
  • Passed HIMA Challenge Test for Sterile Filtration Grade
  • Integrity Tested
  • Integrity testing by bubble point method, pressure collapse method, forward flow method, etc. is possible.
  • Protein adsorption of PES membrane:
  • HSA:0.4μg/cm2
  • Insulin: 2.0 mcg/cm2
  • Gamma globulin: 1.5 mcg/cm2
  • Note)Only for test and research purposes.

Spec

  • Housing: Polypropylene
  • vents: inlet side
  • Membrane Material: Polyethersulfone (PES)
  • Support System: Polypropylene
  • Connection Method: Heat Crimping
  • Voltage: 0.41 MPa (60psi)
  • Filtration direction: If a prefilter is installed, place the prefilter on the inlet side so that the liquid flows in the direction of the arrow.
  • Endotoxin level: LAL test =0.5EU/mL
  • Sterilization: Sterilized/Capsules can be autoclaved at 121°C for 20 minutes (maximum temperature 132°C). Perform the integrity test after autoclaving.
  • Effective Filtration Area: 440 cm2
  • Bubble Point (Water) (Final Membrane): 0.27 MPa (40psi)
  • Pore Diameter (μm): 0.2/0.2
  • Connection Inlet: SB
  • Connection outlet: SB
  • SB= Stepped Hose Port: Can be connected to 6~10 mm (1/4”-3/8”) tubing
  •  

Package size:71×172×71 mm 130 g  [About Package size]

Order No. 62-6178-68
Model No. 6715-3602
Standard price JPY: 18,210 USD: 113.30
Excange rate 1USD= 160.72JPY
Valid price in Japan
Quantity 1piece
Stock in Japan