3-9805-01 Hydrogen Detector NDR-H2-1A

Features

  • This is a hydrogen alarm using a contact combustion type gas sensor made in-house.
  • 3 levels of hydrogen are displayed by LED. (Energization, trouble, gas alarm)

Spec

  • Target gas: hydrogen
  • Size (mm): 100 x 45 x 150
  • Detection Method: Contact Combustion Method
  • Measurement range: 500 to : 20000 ppm (Range error 15%)
  • Warning concentration: 500 ±100 ppm (Standards)
  • Alarm: lamp/light (automatic return), buzzer sound
  • Display: LED (energization, trouble, gas alarm)
  • External output: 1a contact relay (AC100V DC 30 V 0.5 A except resistive load)
  • Operating temperature range: 10 - + 40 ℃ 95 %RH or less (non-condensing)
  • Power supply : AC100V 50/60Hz
  • Weight: about 340g
  •  

Package size:180×260×100 mm 530 g  [About Package size]

Order No. 3-9805-01
Model No. NDR-H2-1A
Standard price JPY: 85,000 USD: 532.82
Excange rate 1USD= 159.53JPY
Valid price in Japan
Quantity 1piece
Stock in Japan
Supplier Stock

Product Catalog

Catalog Name Page
AS ONE Catalog 2025 [Instruments for Laboratory] 625
AS ONE Catalog 2023 [Instruments for Laboratory] 744
AS ONE Catalog 2021 [Instruments for Laboratory] 694
AS ONE Catalog 2019 [Instruments for Laboratory] 672